Atmospheric Measurement Techniques (May 2020)

Single-photon laser-induced fluorescence detection of nitric oxide at sub-parts-per-trillion mixing ratios

  • A. W. Rollins,
  • P. S. Rickly,
  • P. S. Rickly,
  • R.-S. Gao,
  • T. B. Ryerson,
  • S. S. Brown,
  • J. Peischl,
  • J. Peischl,
  • I. Bourgeois,
  • I. Bourgeois

DOI
https://doi.org/10.5194/amt-13-2425-2020
Journal volume & issue
Vol. 13
pp. 2425 – 2439

Abstract

Read online

We describe a newly developed single-photon laser-induced fluorescence sensor for measurements of nitric oxide (NO) in the atmosphere. Rapid tuning of a narrow-band laser on and off of a rotationally resolved NO spectral feature near 215 nm and detection of the red-shifted fluorescence provides for interference-free direct measurements of NO with a detection limit of 1 part per trillion by volume (pptv) for 1 s of integration, or 0.3 pptv for 10 s of integration. Uncertainty in the sensitivity of the instrument is typically ±6–9 %, with no known interferences. Uncertainty in the zero of the detector is shown to be <0.2 pptv. The instrument was deployed on the NASA DC-8 aircraft during the NASA/NOAA FIREX-AQ experiment (Fire Influence on Regional to Global Environments Experiment – Air Quality) during July–September 2019 and provided more than 140 h of NO measurements over 22 flights, demonstrating the ability of this instrument to operate routinely and autonomously. Comparisons with a seasoned chemiluminescence sensor during FIREX-AQ in a variety of chemical environments provides validation and confidence in the accuracy of this technique.