Animals (Dec 2022)

Analytical Validation of the IMMULITE<sup>®</sup> 2000 XPi Progesterone Assay for Quantitative Analysis in Ovine Serum

  • Kristi L. Jones,
  • Ameer A. Megahed,
  • Brittany N. Diehl,
  • Ann M. Chan,
  • Oscar Hernández,
  • Catalina Cabrera,
  • João H. J. Bittar

DOI
https://doi.org/10.3390/ani12243534
Journal volume & issue
Vol. 12, no. 24
p. 3534

Abstract

Read online

Monitoring circulating progesterone (P4) concentration is an important component of basic and applied reproduction research and clinical settings. IMMULITE® 2000 XPi (Siemens) is a newly upgraded fully automated immunoassay system marketed for human use to measure concentrations of different analytes including P4. Our objective was therefore to characterize the analytical performance of the IMMULITE® 2000 XPi P4 immunoassay across the reportable range in ovine serum. This validation of analytical performance included determining (1) linearity, (2) precision through within-run, and between-run coefficient of variation (CV) calculations, (3) accuracy through bias calculations for spiking-recovery bias and interlaboratory (range and average based) bias for two laboratories across the reportable range (0.2–40 ng/mL). The average within-run and between-run precision (CV%) across the reportable range of the IMMULITE® 2000 XPi P4 immunoassay for serum P4 concentration were both ® 2000 XPi P4 immunoassay provides a precise, accurate, reliable, and safe method for measuring P4 concentration ovine serum.