Journal of Cardiovascular Magnetic Resonance (May 2020)

A comparison of cine CMR imaging at 0.55 T and 1.5 T

  • W. Patricia Bandettini,
  • Sujata M. Shanbhag,
  • Christine Mancini,
  • Delaney R. McGuirt,
  • Peter Kellman,
  • Hui Xue,
  • Jennifer L. Henry,
  • Margaret Lowery,
  • Swee Lay Thein,
  • Marcus Y. Chen,
  • Adrienne E. Campbell-Washburn

DOI
https://doi.org/10.1186/s12968-020-00618-y
Journal volume & issue
Vol. 22, no. 1
pp. 1 – 10

Abstract

Read online

Abstract Background There is a renewed interest in lower field magnetic resonance imaging (MRI) systems for cardiovascular magnetic resonance (CMR), due to their favorable physical properties, reduced costs, and increased accessibility to patients with implants. We sought to assess the diagnostic capabilities of high-performance low-field (0.55 T) CMR imaging for quantification of right and left ventricular volumes and systolic function in both healthy subjects and patients referred for clinical CMR. Methods Sixty-five subjects underwent paired exams at 1.5 T using a clinical CMR scanner and using an identical CMR system modified to operate at 0.55 T. Volumetric coverage of the right ventricle (RV) and left ventricles (LV) was obtained using either a breath-held cine balanced steady-state free-precession acquisition or a motion-corrected free-breathing re-binned cine acquisition. Bland-Altman analysis was used to compare LV and RV end-systolic volume (ESV), end-diastolic volume (EDV), ejection fraction (EF), and LV mass. Diagnostic confidence was scored on a Likert-type ordinal scale by blinded readers. Results There were no significant differences in LV and RV EDV between the two scanners (e.g., LVEDV: p = 0.77, bias = 0.40 mL, correlation coefficient = 0.99; RVEDV: p = 0.17, bias = − 1.6 mL, correlation coefficient = 0.98), and regional wall motion abnormality scoring was similar (kappa 0.99). Blood-myocardium contrast-to-noise ratio (CNR) at 0.55 T was 48 ± 7% of the 1.5 T CNR, and contrast was sufficient for endocardial segmentation in all cases. Diagnostic confidence of images was scored as “good” to “excellent” for the two field strengths in the majority of studies. Conclusion A high-performance 0.55 T system offers good bSSFP CMR image quality, and quantification of biventricular volumes and systolic function that is comparable to 1.5 T in patients. Trial registration Clinicaltrials.gov NCT03331380, NCT03581318.

Keywords