Diagnostics (Nov 2022)

T1 Mapping MOLLI 5(3)3 Acquisition Scheme Yields High Accuracy in 1.5 T Cardiac Magnetic Resonance

  • Patrick Krumm,
  • Petros Martirosian,
  • Alexander Brendel,
  • Jens M. Kübler,
  • Jan M. Brendel,
  • Sebastian Gassenmaier,
  • Arne Estler,
  • Meinrad Gawaz,
  • Konstantin Nikolaou,
  • Simon Greulich

DOI
https://doi.org/10.3390/diagnostics12112729
Journal volume & issue
Vol. 12, no. 11
p. 2729

Abstract

Read online

Objectives: To systematically compare two modified Look-Locker inversion recovery (MOLLI) T1 mapping sequences and their impact on (1) myocardial T1 values native, (2) post-contrast and (3) extracellular volume (ECV). Methods: 200 patients were prospectively included for 1.5 T CMR for work-up of ischemic or non-ischemic cardiomyopathies. To determine native and post-contrast T1 for ECV calculation, two different T1 mapping MOLLI acquisition schemes, 5(3)3 (designed for native scans with long T1) and 4(1)3(1)2 (designed for post-contrast scans with short T1), were acquired in identical mid-ventricular short-axis slices. Both schemes were acquired in native and post-contrast scans. Results: Datasets from 163 patients were evaluated (age 55 ± 17 years; 38% female). Myocardial T1 native for 5(3)3 was 1017 ± 42 ms vs. 956 ± 40 ms for 4(1)3(1)2, with mean intraindividual difference −61 ms (p p Conclusions: The T1 MOLLI 5(3)3 acquisition scheme provides a reliable estimation of myocardial T1 for the clinically relevant range of long and short T1 values native and post-contrast. In contrast, the T1 MOLLI 4(1)3(1)2 acquisition scheme may only be used for post-contrast scans according to its designed purpose.

Keywords