International Journal of Infectious Diseases (Jul 2021)

Comparison of Roche and Lumipulse quantitative SARS-CoV-2 antigen test performance using automated systems for the diagnosis of COVID-19

  • Yosuke Hirotsu,
  • Hiroki Sugiura,
  • Makoto Maejima,
  • Miyoko Hayakawa,
  • Hitoshi Mochizuki,
  • Toshiharu Tsutsui,
  • Yumiko Kakizaki,
  • Yoshihiro Miyashita,
  • Masao Omata

Journal volume & issue
Vol. 108
pp. 263 – 269

Abstract

Read online

Background: Severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) continues to spread worldwide. Here, we evaluated the performance of two quantitative antigen (Ag) tests, the Roche and Lumipulse Ag tests, using automated platforms. Methods: We collected 637 nasopharyngeal swab samples from 274 individuals. Samples were subjected to quantitative reverse transcription PCR (RT-qPCR), the Roche Ag test and Lumipulse Ag test. Results: When RT-qPCR was used as a reference, the overall concordance rate of the Roche Ag test was 77.1% (491/637) with 70.0% (341/487) sensitivity and 100% specificity (150/150). When inconclusive results of the Lumipulse Ag test were excluded, the overall concordance rate of the Lumipulse Ag test was 88.3% (467/529) with 84.8% (330/389) sensitivity and 97.9% (137/140) specificity. The overall concordance rate between the Roche and Lumipulse Ag tests was 97.9% (518/529) with 96.7% (322/333) sensitivity and 100% (196/196) specificity. Quantitative Ag levels determined using the Roche and Lumipulse Ag tests were highly correlated (R2 = 0.922). The Roche and Lumipulse Ag tests showed high concordance up to nine days after symptom onset, with progressively lower concordance after that. Conclusions: The Roche and Lumipulse Ag tests showed equivalent assay performance and represent promising approaches for diagnosing coronavirus disease 2019.

Keywords