PLoS ONE (Jan 2014)

Variability of the QuantiFERON®-TB gold in-tube test using automated and manual methods.

  • William C Whitworth,
  • Donald J Goodwin,
  • Laura Racster,
  • Kevin B West,
  • Stella O Chuke,
  • Laura J Daniels,
  • Brandon H Campbell,
  • Jamaria Bohanon,
  • Atheer T Jaffar,
  • Wanzer Drane,
  • Paul A Sjoberg,
  • Gerald H Mazurek

DOI
https://doi.org/10.1371/journal.pone.0086721
Journal volume & issue
Vol. 9, no. 1
p. e86721

Abstract

Read online

BackgroundThe QuantiFERON®-TB Gold In-Tube test (QFT-GIT) detects Mycobacterium tuberculosis (Mtb) infection by measuring release of interferon gamma (IFN-γ) when T-cells (in heparinized whole blood) are stimulated with specific Mtb antigens. The amount of IFN-γ is determined by enzyme-linked immunosorbent assay (ELISA). Automation of the ELISA method may reduce variability. To assess the impact of ELISA automation, we compared QFT-GIT results and variability when ELISAs were performed manually and with automation.MethodsBlood was collected into two sets of QFT-GIT tubes and processed at the same time. For each set, IFN-γ was measured in automated and manual ELISAs. Variability in interpretations and IFN-γ measurements was assessed between automated (A1 vs. A2) and manual (M1 vs. M2) ELISAs. Variability in IFN-γ measurements was also assessed on separate groups stratified by the mean of the four ELISAs.ResultsSubjects (N = 146) had two automated and two manual ELISAs completed. Overall, interpretations were discordant for 16 (11%) subjects. Excluding one subject with indeterminate results, 7 (4.8%) subjects had discordant automated interpretations and 10 (6.9%) subjects had discordant manual interpretations (p = 0.17). Quantitative variability was not uniform; within-subject variability was greater with higher IFN-γ measurements and with manual ELISAs. For subjects with mean TB Responses ±0.25 IU/mL of the 0.35 IU/mL cutoff, the within-subject standard deviation for two manual tests was 0.27 (CI95 = 0.22-0.37) IU/mL vs. 0.09 (CI95 = 0.07-0.12) IU/mL for two automated tests.ConclusionQFT-GIT ELISA automation may reduce variability near the test cutoff. Methodological differences should be considered when interpreting and using IFN-γ release assays (IGRAs).