Diagnostics (May 2022)

Sleep Position Detection with a Wireless Audio-Motion Sensor—A Validation Study

  • Wojciech Kukwa,
  • Tomasz Lis,
  • Jonasz Łaba,
  • Ron B. Mitchell,
  • Marcel Młyńczak

DOI
https://doi.org/10.3390/diagnostics12051195
Journal volume & issue
Vol. 12, no. 5
p. 1195

Abstract

Read online

It is well documented that body position significantly affects breathing indices during sleep in patients with obstructive sleep apnea. They usually worsen while changing from a non-supine to a supine position. Therefore, body position should be an accurately measured and credible parameter in all types of sleep studies. The aim of this study was to specify the accuracy of a neck-based monitoring device (Clebre, Olsztyn, Poland) mounted at the suprasternal notch, in determining a supine and non-supine sleeping position, as well as specific body positions during sleep, in comparison to polysomnography (PSG). A sleep study (PSG along with a neck-based audio-motion sensor) was performed on 89 consecutive patients. The accuracy in determining supine and non-supine positions was 96.9%±3.9% and 97.0%±3.6%, respectively. For lateral positions, the accuracy was 98.6%±2% and 97.4%±4.5% for the right and left side, respectively. The prone position was detected with an accuracy of 97.3%±5.6%. The study showed a high accuracy in detecting supine, as well as other gross positions, during sleep based on a sensor attached to the suprasternal notch, compared to the PSG study. We feel that the suprasternal notch is a promising area for placing wireless sleep study devices.

Keywords