PLoS ONE (Jan 2016)

Rotational Stability of Scaphoid Osteosyntheses: An In Vitro Comparison of Small Fragment Cannulated Screws to Novel Bone Screw Sets.

  • Jochen Erhart,
  • Ewald Unger,
  • Philip Schefzig,
  • Peter Varga,
  • Inga Trulson,
  • Anna Gormasz,
  • Alexander Trulson,
  • Martin Reschl,
  • Michael Hagmann,
  • Vilmos Vecsei,
  • Winfried Mayr

DOI
https://doi.org/10.1371/journal.pone.0156080
Journal volume & issue
Vol. 11, no. 6
p. e0156080

Abstract

Read online

BACKGROUND:The current standard of care for operative repair of scaphoid fractures involves reduction and internal fixation with a single headless compression screw. However, a compression screw in isolation does not necessarily control rotational stability at a fracture or nonunion site. The single screw provides rotational control through friction and bone interdigitation from compression at the fracture site. We hypothesize that osteosyntheses with novel bone screw sets (BSS) equipped with anti-rotational elements provide improved rotational stability. METHODS:Stability of osteosynthesis under increasing cyclic torsional loading was investigated on osteotomized cadaveric scaphoids. Two novel prototype BSS, oblique type (BSS-obl.) and longitudinal type (BSS-long.) were compared to three conventional screws: Acutrak2®mini, HCS®3.0 and Twinfix®. Biomechanical tests were performed on scaphoids from single donors in paired comparison and analyzed by balanced incomplete random block design. Loading was increased by 50 mNm increments with 1,000 cycles per torque level and repeated until a rotational clearance of 10°. Primary outcome measure was the number of cycles to 10° clearance, secondary outcome measure was the maximum rotational clearance for each torque level. FINDINGS:BSS-obl. performed significantly better than Acutrak2®mini and HCS® (p = 0.015, p<0.0001). BSS-long. performed significantly better than HCS® (p = 0.010). No significant difference in performance between BSS-obl. and BSS-long. (p = 0.361), between BSS obl. and Twinfix® (p = 0.50) and BSS long. and Twinfix® (p = 0.667) was detected. Within the torque range up to 200 mNm, four of 21 (19%) BSS-long. and four of 21 (19%) BSS-obl. preparations showed early failure. The same loading led to early failure in four (29%) Twinfix®, seven (50%) Acutrak2®mini and 10 (71%) HCS® of 14 screw samples, respectively. CONCLUSIONS:For both BSS and to a lesser extent for Twinfix® (as dual-component screw), higher rotational stabilities were identified in comparison to single component headless compression screws.