Children (Dec 2022)

Polyvinylpyrrolidone-Coated Catheters Decrease Astrocyte Adhesion and Improve Flow/Pressure Performance in an Invitro Model of Hydrocephalus

  • Leandro Castañeyra-Ruiz,
  • Seunghyun Lee,
  • Alvin Y. Chan,
  • Vaibhavi Shah,
  • Bianca Romero,
  • Jenna Ledbetter,
  • Michael Muhonen

DOI
https://doi.org/10.3390/children10010018
Journal volume & issue
Vol. 10, no. 1
p. 18

Abstract

Read online

The leading cause of ventricular shunt failure in pediatric patients is proximal catheter occlusion. Here, we evaluate various types of shunt catheters to assess in vitro cellular adhesion and obstruction. The following four types of catheters were tested: (1) antibiotic- and barium-impregnated, (2) polyvinylpyrrolidone, (3) barium stripe, and (4) barium impregnated. Catheters were either seeded superficially with astrocyte cells to test cellular adhesion or inoculated with cultured astrocytes into the catheters to test catheter performance under obstruction conditions. Ventricular catheters were placed into a three-dimensional printed phantom ventricular replicating system through which artificial CSF was pumped. Differential pressure sensors were used to measure catheter performance. Polyvinylpyrrolidone catheters had the lowest median cell attachment compared to antibiotic-impregnated (18 cells), barium stripe (17 cells), and barium-impregnated (21.5 cells) catheters after culture (p p < 0.01) catheters. Polyvinylpyrrolidone catheters showed less cellular adhesion and were least likely to be occluded by astrocyte cells. Our findings can help suggest patient-appropriate proximal ventricular catheters for clinical use.