CD4-T cell enumeration in human immunodeficiency virus (HIV)-infected patients: A laboratory performance evaluation of Muse Auto CD4/CD4% system by World Health Organization prequalification of in vitro diagnostics.

  • Ann Ceulemans,
  • Chaymae Bouzahzah,
  • Irena Prat,
  • Willy Urassa,
  • Luc Kestens

DOI
https://doi.org/10.1371/journal.pone.0209677
Journal volume & issue
Vol. 14, no. 1
p. e0209677

Abstract

Read online

BackgroundCD4 T-cell counts are still widely used to assess treatment eligibility and follow-up of HIV-infected patients. The World Health Organization (WHO) prequalification of in vitro diagnostics requested a manufacturer independent laboratory evaluation of the analytical performance at the Institute of Tropical Medicine (ITM) Antwerp, Belgium, of the Muse Auto CD4/CD4% system (Millipore), a new small capillary-flow cytometer dedicated to count absolute CD4-T cells and percentages in venous blood samples from HIV-infected patients.MethodsTwo hundred and fifty (250) patients were recruited from the HIV outpatient clinic at ITM. Accuracy and precision of CD4 T cell counting on fresh EDTA anticoagulated venous blood samples were assessed in the laboratory on a Muse Auto CD4/CD4% system. Extensive precision analyses were performed both on fresh blood and on normal and low stabilized whole blood controls. Accuracy ((bias) was assessed by comparing results from Muse CD4/CD4% to the reference (single-platform FACSCalibur). Clinical misclassification was measured at 500, 350, 200 and 100 cells/μL thresholds.ResultsIntra-assay precision was 90% at all thresholds except at 100 cells/μL (sensitivity = 87%). All samples could be analyzed as there was no repetitive rejection errors recorded.ConclusionsThe Muse Auto CD4/CD4% System performed very well on fresh venous blood samples and met all WHO acceptance criteria for analytical performance of CD4 technologies.