Applied Sciences (Nov 2021)
A Simple Graphene Functionalized Electrochemical Aptasensor for the Sensitive and Selective Detection of Glycated Albumin
Abstract
Glycated albumin (GA) has been previously introduced as a promising biomarker for glycemic monitoring in diabetes patients with thalassemia. In this study, a label-free graphene oxide (GO)-modified aptasensor was developed for the rapid detection of GA. The fabrication of the aptasensor was dependent on the covalent interaction of the amine-functionalized GA-specific aptamer with the carboxylic groups of GO. Square wave voltammetry (SWV) analysis was carried out for the measurement of GA-aptamer binding to their specific proteins. The peak current changes before and after incubation with GA protein were directly proportional to the concentration. The developed aptasensor exhibited a broad linearity (1–10,000 µg mL−1), a low detection limit (LOD) of 0.031 µg mL−1, and high selectivity for GA detection. In addition, the aptasensor was successfully applied to detect GA in both spiked and clinical serum samples. The comparison of the developed method with a commercial assay validated the reliability of the aptasensor for clinical application. Therefore, the newly developed aptasensor is a promising tool for GA measurements in diabetic patients with underlying thalassemia.