Shiyan dongwu yu bijiao yixue (Jun 2024)

Effects of 1 470 nm Semiconductor Laser on Vaporization Ablation, Cutting, and Coagulation in Ex Vivo Animal Tissue

  • ZHENG Guo,
  • PAN Yongming,
  • HUANG Junjie,
  • ZHANG Hui,
  • YU Chen,
  • CHEN Minli,
  • XU Qingfeng,
  • HUANG Heng

DOI
https://doi.org/10.12300/j.issn.1674-5817.2023.161
Journal volume & issue
Vol. 44, no. 3
pp. 279 – 288

Abstract

Read online

Objective To observe the effects of a 1 470 nm semiconductor laser on vaporization cutting, coagulation, and thermal injury of ex vivo animal tissues, aiming to explore the feasibility of its application in the treatment of benign prostatic hyperplasia. Methods The experimental group and control group were treated with HANS-D1 and ML-DD01FI 1 470 nm semiconductor laser therapy equipment, respectively. Fresh ex vivo pig bladder tissue was exposed to lasers with the optical fiber placed at distances of 0.5 cm and 1 cm from the tissue for 5 s. The effects of layers at powers of 60, 90, 120, 150, and 160 W on tissue injury were observed. Ex vivo dog prostate and pig kidney tissues were used for vaporization ablation and cutting to observe the effects of lasers at the same power levels on tissue vaporization and cutting thermal injury. Additionally, in coagulation mode, the effects of 30, 40, and 50 W semiconductor lasers on tissue coagulation were observed after irradiating ex vivo pig kidney tissue for 5, 10, and 15 seconds.Results When the optical fiber was placed 1 cm away from the tissue, the 1 470 nm semiconductor lasers did not cause accidental damage to adjacent normal bladder tissue. However, at a distance of 0.5 cm, the 120 W, 150 W, or 160 W lasers caused slight damage to the bladder tissue. In addition, with the increase in output power, the vaporization ablation efficiency of 60-160 W lasers on dog prostate tissue gradually increased, showing a good linear correlation between vaporization volume and total energy consumption (P0.05). Moreover, when performing vaporization cutting on ex vivo pig kidney tissue with a cross-sectional area of 1 cm2, the efficiency of vaporization cutting by the 60-160 W 1 470 nm semiconductor lasers increased with the increase in output power (P0.05). Conclusion The 1 470 nm semiconductor laser shows good vaporization ablation, cutting, and coagulation effect on ex vivo tissues, with a good linear correlation between the effect and the output energy.

Keywords