Zhongguo shuxue zazhi (Nov 2023)

Performance verification experiment for enzyme-linked immunosorbent assay in blood screening laboratory

  • Humin LIU,
  • Xue CHEN,
  • Wenjun LI,
  • Yinglan LUO,
  • Jing MEI

DOI
https://doi.org/10.13303/j.cjbt.issn.1004-549x.2023.11.019
Journal volume & issue
Vol. 36, no. 11
pp. 1049 – 1054

Abstract

Read online

Objective To verificate the performance of enzyme-linked immunosorbent assay (ELISA) in blood screening laboratory. Methods The repeatability, precision, sensitivity, specificity, compliance, detection limit and anti-interference of ELISA items in the laboratory detection system were verified. Results The repeatability was 100%.The intra batch imprecision of each system was less than 10%, and the inter batch imprecision was less than 15%. The sensitivity, specificity and compliance were 100%, with the minimum detection limits of the two reagents at 0.75 NCU/mL and 0.25 NCU/mL respectively, The anti-interference met the requirements of the reagent manual. Conclusion The analysis of the performance verification data of ELISA test items will help continuously improve the performance of detection system and ensure the safety of clinical blood use.