Revista do Colégio Brasileiro de Cirurgiões (Feb 2018)
Glucose level evaluation in monopedicled rectus abdominis myocutaneous flap after venous occlusion: experimental study in rats
Abstract
ABSTRACT Objective: to validate an experimental model for the measurement of glycemic levels in surgical flaps with the use of common glucometers, and to analyze the diagnostic criteria for hypoperfusion of such flaps. Methods: we performed vertical myocutaneous rectus abdominis flaps with upper pedicles bilaterally in 20 male Wistar rats, divided into two groups: with and without venous occlusion of the pedicle. We measured glucose levels in the flaps and in the systemic circulation with standard glucometers. We tested the accuracy of alternative diagnostic criteria for the detection of hypoperfusion. Results: from 15 minutes of venous occlusion on, there was a significant reduction in glucose levels measured in the congested flap (p<0.001). Using a minimum difference of 20mg/dl in the glycemic levels between the flap and systemic blood, 30 minutes after occlusion, as a diagnostic criterion, the sensitivity was 100% (95% CI 83.99-100%) and specificity of 90% (95% CI 69.90-97.21%) for the diagnosis of flap congestion. Conclusion: It is possible to measure glucose levels in vertical myocutaneous rectus abdominis flaps of Wistar rats, perfused or congested, using a common glucometer. The diagnostic criteria that compare the glucose levels in the flaps with the systemic ones were more accurate in the evaluation of tissue perfusion.