The Journal of Engineering (Oct 2019)

Monte–Carlo simulation of wall correction factor of graphite cavity ionisation chamber

  • Bo Wang,
  • Jinjie Wu,
  • Haiyan Du,
  • Siming Guo,
  • Shiwei Ren

DOI
https://doi.org/10.1049/joe.2018.9078

Abstract

Read online

The purpose of this article is to discuss the method of determining the wall correction factor of graphite cavity ionisation chamber. The method is the ‘equal effect wall thickness’ simulation and direct calculation of the wall correction factor of the std10cm^3 and std30cm^3 spherical graphite cavity ionisation chamber in the national air-specific kinetic energy reference group (National Institute of Standards and Technology NIST), respectively, in the NIST10cm^3 and NIST30cm^3 spherical graphite cavity ionisation chamber. The calculated results are in agreement with the values released by NIST in 0.1%. Conclusions of the equivalent wall thickness simulation method and the direct calculation simulation method meet each other's requirements, which provide a new idea for the determination of wall correction factor.