PLoS ONE (Jan 2018)

Comparison of strain elastography, point shear wave elastography using acoustic radiation force impulse imaging and 2D-shear wave elastography for the differentiation of thyroid nodules.

  • Georgia Kyriakidou,
  • Mireen Friedrich-Rust,
  • Dimitra Bon,
  • Ishani Sircar,
  • Christopher Schrecker,
  • Dimitra Bogdanou,
  • Eva Herrmann,
  • Joerg Bojunga

DOI
https://doi.org/10.1371/journal.pone.0204095
Journal volume & issue
Vol. 13, no. 9
p. e0204095

Abstract

Read online

PURPOSE:The aim of the study was to compare three different elastography methods, namely Strain Elastography (SE), Point Shear-Wave Elastography (pSWE) using Acoustic Radiation Force Impulse (ARFI)-Imaging and 2D-Shear Wave Elastography (2D-SWE), in the same study population for the differentiation of thyroid nodules. MATERIALS AND METHODS:All patients received a conventional ultrasound scan, SE and 2D-SWE, and all patients except for two received ARFI-Imaging. Cytology/histology of thyroid nodules was used as a reference method. SE measures the relative stiffness within the region of interest (ROI) using the surrounding tissue as reference tissue. ARFI mechanically excites the tissue at the ROI using acoustic pulses to generate localized tissue displacements. 2D-SWE measures tissue elasticity using the velocity of many shear waves as they propagate through the tissue. RESULTS:84 nodules (73 benign and 11 malignant) in 62 patients were analyzed. Sensitivity, specificity and NPV of SE were 73%, 70% and 94%, respectively. Sensitivity, specificity and NPV of ARFI and 2D-SWE were 90%, 79%, 98% and 73%, 67%, 94% respectively, using a cut-off value of 1.98m/s for ARFI and 2.65m/s (21.07kPa) for 2D-SWE. The AUROC (Area under the Receiver Operating Characteristic) of SE, ARFI and 2D-SWE for the diagnosis of malignant thyroid nodules were 52%, 86% and 71%, respectively. A significant difference in AUROC was found between SE and ARFI (p = 0.008), while no significant difference was found between ARFI and SWE (86% vs. 71%, p = 0.31), or SWE and SE (71% vs. 52%, p = 0.26). CONCLUSION:pSWE using ARFI and 2D-SWE showed comparable results for the differentiation of thyroid nodules. ARFI was superior to elastography using SE.