Materials (May 2023)

Bacterial Detection, Deformation, and Torque Loss on Dental Implants with Different Tapered Connections Compared with External Hexagon Connection after Thermomechanical Cycling

  • Anselmo Agostinho Simionato,
  • Emerson de Souza Santos,
  • Adriana Cláudia Lapria Faria,
  • Cássio do Nascimento,
  • Renata Cristina Silveira Rodrigues,
  • Ricardo Faria Ribeiro

DOI
https://doi.org/10.3390/ma16113904
Journal volume & issue
Vol. 16, no. 11
p. 3904

Abstract

Read online

The relationship between bacterial infiltration and internal conical Implant-Abutment Interfaces (IAIs) with different conicities still requires investigations that can offer valuable information in the clinical understanding of peri-implant health. The present study aimed to verify the bacterial infiltration of two internal conical connections with an angulation of 11.5° and 16° with the external hexagonal connection as a comparative after thermomechanical cycling using saliva as a contaminant. Test (n = 10) and control (n = 3) groups were set up. Evaluations were made on torque loss, Scanning Electron Microscopy (SEM), and Micro Computerized Tomography (MicroCT) after performing 2 × 106 mechanical cycles (120 N) and 600 thermal cycles (5°–55° C) with 2 mm lateral displacement. The contents of the IAI were collected for microbiological analysis. There was a difference (p p < 0.05). In conclusion, the IAI environment may favor a microbiological profile different from that of saliva and the thermocycling condition may alter the microbial profile found in the IAI.