Indian Journal of Ophthalmology (Jan 2021)

Saving space: Comparing mini - logMAR with standard logMAR visual acuity

  • Farnaz Kauser,
  • Abadan K Amitava,
  • Juhi Saxena,
  • S Aisha Raza,
  • Anam Masood,
  • Md Shahid Alam

DOI
https://doi.org/10.4103/ijo.IJO_2391_19
Journal volume & issue
Vol. 69, no. 1
pp. 48 – 51

Abstract

Read online

Purpose: Assessing visual acuity (VA) is the cornerstone of an ophthalmic workup and needs VA charts in a four or six meters space. The objective of this study was to compare the performance of distant VA (DVA) on one meter mini-logMAR (MLM) with a standard six meter logMAR (SLM) chart. Methods: We developed a MLM chart to be used at 1 m with +1.0 D spectacles, by reducing the SLM chart designed for 6 meters, to 1/6th its size, using AutoCAD version 2014. On an initial cohort, we obtained DVA on the two charts by optometrist trainees, masked to the outcomes on the different tests. We performed regression and checked agreement between the two measurements. Subsequently, on a new cohort, we validated the performance of the MLM. Results: Of the 56 initial subjects, mean DVA with SLM was 0.44 ± 0.13 and with MLM was 0.45 ± 0.13; mean difference of -0.01 ± 0.02, 95%CI: 0.007 to 0.018; P < 0.0001 on paired t-test. There was a significant correlation: r = 0.99; r2 = 0.98, P < 0.0001. On an average, DVA with MLM was less than a letter worse than with SLM. The regression formula obtained: SLM DVA = -0.1312 + 1.0014 x (MLM DVA). The validation study revealed no significant difference (P = 0.29) between the predicted standard DVA calculated by the regression formula and the actual standard. Conclusion: We suggest that we can deduce distance logMAR VA from a mini-logMAR chart as devised and used by us. This will take less space, be portable and allow congenial interaction with patients.

Keywords