Effect of chemical mechanical polishing on surface nature of titanium implants FT-IR and wettability data of titanium implants surface after chemical mechanical polishing implementation
Abstract
Bioactivity of titanium depends on the quality and characteristics of its surface oxide film. Through implementation of chemical mechanical polishing (CMP) process on titanium plates, a protective oxide (titania) film grows on the titanium based implant surface. In this article, surface properties of the titanium oxide are investigated as a function of the CMP process conditions. Surface responses were evaluated in terms of wettability, nano-scale surface roughness and material removal rates (MRR). Surface chemical compositions were also evaluated through Fourier transform infrared spectroscopy (FT-IR).