Zhongliu Fangzhi Yanjiu (Nov 2023)
Application of Linear Accelerator on Boarding Kilovolt Fan Beam CT and Megavolt Cone Beam CT System on Set-up Errors During Radiation Treatment
Abstract
Objective To quantify the setup errors for the different anatomical sites of patients who received intensity-modulated radiotherapy (IMRT) with linear accelerator on-board kilovolt fan beam CT(kV-FBCT) as non-isocenter IGRT and megavolt cone beam CT (MV-CBCT) as isocenter IGRT. Methods A retrospective analysis was performedon 70 patients who underwent radiotherapy, kV-FBCT, and/or MV-CBCT scans after each routine setup prior to IMRT. The average displacement (M), systematic error (Σ), and random error (б) at different treatment sites in the left-right, anterior-posterior, and cranial-caudal directions were calculated according to the individual displacements. The formula 2.5Σ+0.7б was used to estimate the PTV margin in respective direction. For each single patient, the root mean square in three directions was used as 3D displacement. Results A total of 1130 displacements were recorded in the 70 patients. The PTV margin was estimated to be 1.9-3.1 mm in head and neck cancer, 2.8-5.1 mm in thoracic cancer, 4.6-5.1 mm in breast cancer, 3.0-5.5 mm in upper abdominal cancer, and 3.5-6.8 mm in pelvic tumor. For the 3D mean displacements, the head and neck, thoracic, breast, upper abdominal, and pelvic cancer were 2.4±1.0, 4.0±1.6, 4.1±2.0, 4.6±2.1, and 4.6±2.1 mm, respectively. The average 3D displacement obtained by kV-FBCT and MV-CBCT were 4.1 and 3.4 mm, respectively (P=0.212). Conclusion The quantitative setup-error data can be obtained using linear accelerator on-board FBCT, and the non-isocenter IGRT induced set-up error cannot be negligible.