Fisioterapia em Movimento ()

The use of a sphygmomanometer to measure shoulder isometric strength: a validity and reliability study

  • Liam Anthony Toohey,
  • Marcos de Noronha,
  • Guilherme S Nunes

DOI
https://doi.org/10.1590/1980-5918.030.003.ao17
Journal volume & issue
Vol. 30, no. 3
pp. 587 – 593

Abstract

Read online

Abstract Introduction: A sphygmomanometer is an instrument commonly used to measure blood pressure that can potentially be used to objectively assess shoulder isometric muscle strength. Objective: To establish the criterion validity and the intra-rater reliability of the sphygmomanometer for the assessment of shoulder isometric muscular strength compared to the handheld dynamometer. To determine if there is a statistically significant difference for shoulder strength between dominant and non-dominant sides. Methods: A test-retest study design was developed, where a rater assessed shoulder flexion and abduction isometric strength of 13 healthy university students, using a commercially available sphygmomanometer and a handheld dynamometer. Results: The criterion validity of the sphygmomanometer was found to be good for both right and left shoulder flexion and abduction strength assessment (Pearson’s r = 0.90-0.97). The intra-rater reliability of the sphygmomanometer was calculated to be good for both right and left flexion and abduction (ICC = 0.96-0.99). The handheld dynamometer also showed good intra-rater reliability for each of the strength measures assessed (ICC = 0.94-0.98). Significant differences (p < 0.01) were identified between dominant and non-dominant sides for shoulder strength. Conclusion: A sphygmomanometer is a simple and easily accessible tool that provides clinicians with accurate objective values for isometric shoulder strength assessment.

Keywords