Guoji Yanke Zazhi (Dec 2013)

Comparison of the prediction error in cataract surgery with Lenstar and conventional ultrasound

  • Hou-Cheng Liang,
  • Ting Ma,
  • Tan Long

DOI
https://doi.org/10.3980/j.issn.1672-5123.2013.12.09
Journal volume & issue
Vol. 13, no. 12
pp. 2396 – 2399

Abstract

Read online

AIM: To compare the prediction errors(PE)in cataract surgery with Lenstar and conventional ultrasound. METHODS: The data of age-related cataract patients were retrospectively analyzed from March, 2013 to June, 2013 in our hospital. Preoperative measurements of ocular biological parameters and calculation of intraocular lens(IOL)degree using SRK/T's formula with ultrasound, keratometry and Lenstar were performed. Cataract extraction combined with IOL implantation in capsule was taken in every patient. Retinoscopy was taken postoperatively after 3 months. Comparison of the two inspection methods for measuring axial length, mean corneal curvature and postoperative refractive PE and absolute value of PE(APE). RESULTS: Preoperative axial length was 24.68±1.70mm and 24.42±1.65mm with Lenstar and ultrasound, respectively, and there was significant difference(t=-12.688, Pr=0.992, Pt=-1.241, P=0.217), but was the significant correlation(r=0.963, Pt=-5.494, Pt=6.379, PCONCLUSION: Accurate ocular biological parameters can be achieved with Lenstar, and postoperative PE is more precise with Lenstar compared with conventional ultrasound. Lenstar can be used for precise calculation of IOL degree in cataract operation.