PLoS ONE (Jan 2022)
Effect of scanning-aid agents on the scanning accuracy in specially designed metallic models: A laboratory study.
Abstract
The advent of intraoral scanning methods has caused a paradigm shift in dentistry. However, despite their many advantages, intraoral scanners cannot accurately recognize the metallic surfaces of prothesis. Therefore, this experiment was designed to verify the effect of scanning-aid agents on the scanning accuracy using metallic reference models. Three different types of metallic reference models (inlay, onlay, and bridge) were specially designed and produced using a milling machine to simulate intraoral dental restorations. Three experimental groups (application of ScanCure, IP Scan Spray, and VITA Powder Scan Spray) were set up and scanned images (each n = 5) were acquired using the I500® intraoral scanner. The reference datasets were established by a 3D design that reflected the deviations between the measured distances and previously planned distances on the reference models. All acquired experimental datasets were digitally superimposed and compared with the reference datasets. Intragroup comparisons (precision, n = 10) were also performed. The root mean square (RMS) values of trueness in the ScanCure and IP groups were significantly more accurate than those of the VITA group in the inlay and onlay reference models (p < 0.05). Notably, in the bridge reference model, the liquid-type ScanCure group showed the highest accuracy of trueness, with statistical significance (p < 0.05). However, the RMS values of precision were not significantly different among the groups. These findings suggest that liquid-type scanning agents can be effectively used to obtain more accurate scan images of intraoral metallic dental restorations.