Biomechanical Analysis of the Efficacy of Locking Plates during Cyclic Loading in Metacarpal Fractures

  • Stefanie Doht,
  • Rainer H. Meffert,
  • Michael J. Raschke,
  • Torsten Blunk,
  • Sabine Ochman

DOI
https://doi.org/10.1155/2014/648787
Journal volume & issue
Vol. 2014

Abstract

Read online

Purpose. To analyse the biomechanical characteristics of locking plates under cyclic loading compared to a nonlocking plate in a diaphyseal metacarpal fracture. Methods. Oblique diaphyseal shaft fractures in porcine metacarpal bones were created in a biomechanical fracture model. An anatomical reduction and stabilization with a nonlocking and a comparable locking plate in mono- or bicortical screw fixation followed. Under cyclic loading, the displacement, and in subsequent load-to-failure tests, the maximum load and stiffness were measured. Results. For the monocortical screw fixation of the locking plate, a similar displacement, maximum load, and stiffness could be demonstrated compared to the bicortical screw fixation of the nonlocking plate. Conclusions. Locking plates in monocortical configuration may function as a useful alternative to the currently common treatment with bicortical fixations. Thereby, irritation of the flexor tendons would be avoided without compromising the stability, thus enabling the necessary early functional rehabilitation.