Sensors (Jun 2022)

Development of Joint Activity Angle Measurement and Cloud Data Storage System

  • Chiu-Ching Tuan,
  • Yi-Chao Wu,
  • Wen-Ling Yeh,
  • Chun-Chieh Wang,
  • Chi-Heng Lu,
  • Shao-Wei Wang,
  • Jack Yang,
  • Tsair-Fwu Lee,
  • Hsuan-Kai Kao

DOI
https://doi.org/10.3390/s22134684
Journal volume & issue
Vol. 22, no. 13
p. 4684

Abstract

Read online

In this study, we developed a range of motion sensing system (ROMSS) to simulate the function of the elbow joint, with errors less than 0.76 degrees and 0.87 degrees in static and dynamic verification by the swinging and angle recognition modules, respectively. In the simulation process, the ɣ correlation coefficient of the Pearson difference between the ROMSS and the universal goniometer was 0.90, the standard deviations of the general goniometer measurements were between ±2 degrees and ±2.6 degrees, and the standard deviations between the ROMSS measurements were between ±0.5 degrees and ±1.6 degrees. With the ROMSS, a cloud database was also established; the data measured by the sensor could be uploaded to the cloud database in real-time to provide timely patient information for healthcare professionals. We also developed a mobile app for smartphones to enable patients and healthcare providers to easily trace the data in real-time. Historical data sets with joint activity angles could be retrieved to observe the progress or effectiveness of disease recovery so the quality of care could be properly assessed and maintained.

Keywords