BMC Ophthalmology (Feb 2022)

Clinical outcomes of a beveled tip, ultra-high speed, 25-gauge pars plana vitrectomy system

  • Harvey S. Uy,
  • Vicente Lorenzo O. Cabahug,
  • Jose Carlo M. Artiaga,
  • Pik Sha Chan,
  • Jordan T. Famadico

DOI
https://doi.org/10.1186/s12886-022-02311-3
Journal volume & issue
Vol. 22, no. 1
pp. 1 – 11

Abstract

Read online

Abstract Objective To report the clinical outcomes of a 25-gauge, beveled-tip, 10,000 cuts-per-minute (cpm) microincisional vitrectomy surgery (MIVS) system. Methods Prospective case series of eyes undergoing primary pars plana vitrectomy (PPV) for common vitreoretinal indications. Main outcome measures were: rate of achieving surgical objectives, operative times, number of surgical steps, use of ancillary instruments, corrected distance visual acuity (CDVA), and adverse events (AE). Results The surgical objectives were achieved in all eyes. Mean total operative time (TOT), core, shave and total vitrectomy times were 1891 ± 890, 204 ± 120, 330 ± 320, 534 ± 389 s, respectively. Mean number of surgical steps was 4.3 ± 1.5. Mean number of ancillary instruments used was 4.5 ± 1.9. Mean CDVA improved by 0.53 ± 0.56 logMAR units (P < 0.001) 3 months postoperatively. AE included elevated IOP (8%), hypotony (6%), and re-detachment (2%). Majority (82%) had no postoperative discomfort. The number of surgical steps demonstrated a positive correlation with TOT (p < 0.05), number of ancillary instruments used (p < 0.05), and postoperative Day 1 IOP (p < 0.05). The number of times ancillary instrumentation was used demonstrated a positive correlation with TOT (p < 0.05). Conclusion Beveled-tip, 10,000 cpm MIVS system effectively and safely performs common VR procedures of varying complexity and may reduce operative times and use of ancillary instrumentation.

Keywords