PLoS ONE (Jan 2019)
Thermodilution vs estimated Fick cardiac output measurement in an elderly cohort of patients: A single-centre experience.
Abstract
AimsPatients referred to the cath-lab are an increasingly elderly population. Thermodilution (TD, gold standard) and the estimated Fick method (eFM) are interchangeably used in the clinical routine to measure cardiac output (CO). However, their correlation in an elderly cohort of cardiac patients has not been tested so far.MethodsA single, clinically-indicated right heart catheterization was performed on each patient with CO estimated by eFM and TD in 155 consecutive patients (75.1±6.8 years, 57.7% male) between April 2015 and August 2017. Whole Body Oxygen Consumption (VO2) was assumed by applying the formulas of LaFarge (LaF), Dehmer (De) and Bergstra (Be). CO was indexed to body surface area (Cardiac Index, CI).ResultsCI-TD showed an overall moderate correlation to CI-eFM as assessed by LaF, De or Be (r2 = 0.53, r2 = 0.54, r2 = 0.57, all p ConclusionAlthough CI-eFM shows an overall reasonable correlation with CI-TD, the predictive value in a single patient is low. CI-eFM cannot replace CI-TD in elderly patients.