Ophthalmology Science (May 2024)
Spectral-Domain and Swept-Source OCT Angiographic Scans Yield Similar Drusen Measurements When Processed with the Same Algorithm
Abstract
Purpose: An algorithm developed to obtain drusen area and volume measurements using swept-source OCT angiography (SS-OCTA) scans was tested on spectral-domain OCT angiography (SD-OCTA) scans. Design: Retrospective study. Participants: Forty pairs of scans from 27 eyes with intermediate age-related macular degeneration and drusen. Methods: Patients underwent both SD-OCTA and SS-OCTA imaging at the same visit using the 6 mm × 6 mm OCTA scan patterns. Using the same algorithm, we obtained drusen area and volume measurements within both 3 mm and 5 mm fovea-centered circles. Paired 2-sample t-tests were performed along with Pearson’s correlation tests. Main Outcome Measures: Mean square root (sqrt) drusen area and cube root (cbrt) drusen volume within the 3 mm and 5 mm fovea-centered circles. Results: Mean sqrt drusen area values from SD-OCTA and SS-OCTA scans were 1.57 (standard deviation [SD] 0.57) mm and 1.49 (SD 0.58) mm in the 3 mm circle and 1.88 (SD 0.59) mm and 1.76 (SD 0.58) mm in the 5 mm circle, respectively. Mean cbrt drusen volume measurements were 0.54 (SD 0.19) mm and 0.51 (SD 0.20) mm in the 3 mm circle, and 0.60 (SD 0.17) mm and 0.57 (SD 0.17) mm in the 5 mm circle. Small differences in area and volume measurements were found (all P 0.97; all P < 0.001). Conclusions: An algorithm originally developed for SS-OCTA scans performs well when used to obtain drusen volume and area measurements from SD-OCTA scans; thus, a separate SD-OCT structural scan is unnecessary to obtain measurements of drusen. Financial Disclosure(s): Proprietary or commercial disclosure may be found in the Footnotes and Disclosures at the end of this article.