Nukleonika (Dec 2020)

Efficient reading of thermoluminescent dosimeter signals using semiconductor detectors

  • Sobotka Piotr,
  • Kliś Bartłomiej,
  • Baranowska Zuzanna,
  • Wołoszczuk Katarzyna,
  • Rutkowska Katarzyna,
  • Woliński Tomasz

DOI
https://doi.org/10.2478/nuka-2020-0034
Journal volume & issue
Vol. 65, no. 4
pp. 223 – 227

Abstract

Read online

The aim of this experimental work was to examine whether semiconductor photodetectors may be applied for the efficient reading of thermoluminescent dosimeter (TLD) signals. For this purpose, a series of experiments have been performed at the Department of Physics, Warsaw University of Technology, in cooperation with the Central Laboratory for Radiological Protection (CLOR). Specifically, the measurement system proposed here has been designed to detect a signal from TLDs that use a semiconductor detector operating in conditions analogous to those met when using commercial devices equipped with a classic photomultiplier. For the experimental tests, the TLDs were irradiated with a beam of 137Cs radiation in the accredited Laboratory for Calibration of Dosimetric and Radon Instruments. Eventually, a comparison of the results obtained with a semiconductor detector (ID120) and a commercial TLD reader with a photomultiplier tube (RADOS) were made.

Keywords