Scientific Reports (Jul 2017)

Multi-pinhole fluorescent x-ray computed tomography for molecular imaging

  • Tenta Sasaya,
  • Naoki Sunaguchi,
  • Kazuyuki Hyodo,
  • Tsutomu Zeniya,
  • Tetsuya Yuasa

DOI
https://doi.org/10.1038/s41598-017-05179-2
Journal volume & issue
Vol. 7, no. 1
pp. 1 – 12

Abstract

Read online

Abstract We propose a multi-pinhole fluorescent x-ray computed tomography (mp-FXCT) technique for preclinical molecular imaging that can provide the complete data necessary to produce 3-D tomographic images during anaesthesia. In this method, multiple projections are simultaneously acquired through a multi-pinhole collimator with a 2-D detector and full-field volumetric beam to accelerate the data acquisition process and enhance the signal-to-noise ratios of the projections. We constructed a 15-pinhole mp-FXCT imaging system at beamline ARNE-7A at KEK and performed preliminary experiments to investigate its imaging properties using physical phantoms and a non-radioactive I imaging agent. The mp-FXCT system could detect an I concentration of 0.038 mg/ml, the minimum required for in-vivo imaging, at a spatial resolution of about 0.3 mm during a data acquisition time of 90 min, which is less than the time for which anaesthesia is effective and suggests that preclinical molecular imaging is feasible with mp-FXCT.