Diagnostics (Aug 2020)

Deep Learning Versus Iterative Reconstruction for CT Pulmonary Angiography in the Emergency Setting: Improved Image Quality and Reduced Radiation Dose

  • Marc Lenfant,
  • Olivier Chevallier,
  • Pierre-Olivier Comby,
  • Grégory Secco,
  • Karim Haioun,
  • Frédéric Ricolfi,
  • Brivaël Lemogne,
  • Romaric Loffroy

DOI
https://doi.org/10.3390/diagnostics10080558
Journal volume & issue
Vol. 10, no. 8
p. 558

Abstract

Read online

To compare image quality and the radiation dose of computed tomography pulmonary angiography (CTPA) subjected to the first deep learning-based image reconstruction (DLR) (50%) algorithm, with images subjected to the hybrid-iterative reconstruction (IR) technique (50%). One hundred forty patients who underwent CTPA for suspected pulmonary embolism (PE) between 2018 and 2019 were retrospectively reviewed. Image quality was assessed quantitatively (image noise, signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR)) and qualitatively (on a 5-point scale). Radiation dose parameters (CT dose index, CTDIvol; and dose-length product, DLP) were also recorded. Ninety-three patients were finally analyzed, 48 with hybrid-IR and 45 with DLR images. The image noise was significantly lower and the SNR (24.4 ± 5.9 vs. 20.7 ± 6.1) and CNR (21.8 ± 5.8 vs. 18.6 ± 6.0) were significantly higher on DLR than hybrid-IR images (p p vol (4.8 ± 1.4 vs. 4.0 ± 1.2 mGy) and DLP (157.9 ± 44.9 vs. 130.8 ± 41.2 mGy∙cm) were lower on DLR than hybrid-IR images. DLR both significantly improved the image quality and reduced the radiation dose of CTPA examinations as compared to the hybrid-IR technique.

Keywords