Journal of Cardiovascular Development and Disease (Feb 2024)

Early Feasibility Study of a Hybrid Tissue-Engineered Mitral Valve in an Ovine Model

  • Ramin Zareian,
  • Samuel D. Zuke,
  • Daisuke Morisawa,
  • Roger S. Geertsema,
  • Mariwan Majid,
  • Clinton Wynne,
  • Jeffrey C. Milliken,
  • Arash Kheradvar

DOI
https://doi.org/10.3390/jcdd11020069
Journal volume & issue
Vol. 11, no. 2
p. 69

Abstract

Read online

Tissue engineering aims to overcome the current limitations of heart valves by providing a viable alternative using living tissue. Nevertheless, the valves constructed from either decellularized xenogeneic or purely biologic scaffolds are unable to withstand the hemodynamic loads, particularly in the left ventricle. To address this, we have been developing a hybrid tissue-engineered heart valve (H-TEHV) concept consisting of a nondegradable elastomeric scaffold enclosed in a valve-like living tissue constructed from autologous cells. We developed a 21 mm mitral valve scaffold for implantation in an ovine model. Smooth muscle cells/fibroblasts and endothelial cells were extracted, isolated, and expanded from the animal’s jugular vein. Next, the scaffold underwent a sequential coating with the sorted cells mixed with collagen type I. The resulting H-TEHV was then implanted into the mitral position of the same sheep through open-heart surgery. Echocardiography scans following the procedure revealed an acceptable valve performance, with no signs of regurgitation. The valve orifice area, measured by planimetry, was 2.9 cm2, the ejection fraction reached 67%, and the mean transmitral pressure gradient was measured at 8.39 mmHg. The animal successfully recovered from anesthesia and was transferred to the vivarium. Upon autopsy, the examination confirmed the integrity of the H-TEHV, with no evidence of tissue dehiscence. The preliminary results from the animal implantation suggest the feasibility of the H-TEHV.

Keywords