Scientific Reports (Jun 2021)

Assessment of reliability and validity of the 5-scale grading system of the point-of-care immunoassay for tear matrix metalloproteinase-9

  • Minjeong Kim,
  • Ja Young Oh,
  • Seon Ha Bae,
  • Seung Hyeun Lee,
  • Won Jun Lee,
  • Yeoun Sook Chun,
  • Kyoung Woo Kim

DOI
https://doi.org/10.1038/s41598-021-92020-6
Journal volume & issue
Vol. 11, no. 1
pp. 1 – 9

Abstract

Read online

Abstract We evaluated the reliability and validity of the 5-scale grading system to interpret the point-of-care immunoassay for tear matrix metalloproteinase (MMP)-9. Six observers graded red bands of photographs of the readout window in MMP-9 immunoassay kit (InflammaDry) two times with 2-week interval based on the 5-scale grading system (i.e. grade 0–4). Interobserver and intraobserver reliability were evaluated using intraclass correlation coefficients. The interobserver agreements were analyzed according to the severity of tear MMP-9 expression. To validate the system, a concentration calibration curve was made using MMP-9 solutions with reference concentrations, then the distribution of MMP-9 concentrations was analyzed according to the 5-scale grading system. Both intraobserver and interobserver reliability was excellent. The readout grades were significantly correlated with the quantified colorimetric densities. The interobserver variance of readout grades had no correlation with the severity of the measured densities. The band density continued to increase up to a maximal concentration (i.e. 5000 ng/mL) according to the calibration curve. The difference of grades reflected the change of MMP-9 concentrations sensitively, especially between grade 2 and 4. Together, our data indicate that the subjective 5-scale grading system in the point-of-care MMP-9 immunoassay is an easy and reliable method with acceptable accuracy.