A Transit Time-Resolved Microflow Cytometry-Based Agglutination Immunoassay for On-Site C-Reactive Protein Detection

  • Jianxi Qu,
  • Yushan Zhang,
  • Mathieu Chenier,
  • Chang-qing Xu,
  • Lan Chen,
  • Yonghong Wan

DOI
https://doi.org/10.3390/mi12020109
Journal volume & issue
Vol. 12, no. 2
p. 109

Abstract

Read online

An accurate and rapid microflow cytometry-based agglutination immunoassay (MCIA) suitable for on-site antibody or antigen detection was proposed. In this study, quantitative C-reactive protein (CRP) detection was chosen as a model assay in order to demonstrate the detection principle. The average transit time was employed to estimate the extent of the agglutination reaction and improve the detection accuracy as compared to the intensity-dependent methods. The detection time was less than 8 min. and only a 20 µL serum sample was needed for each test. The results showed a linear relationship between the average transit time of aggregates and CRP concentrations ranging from 0 to 1 µg/mL. The R2 of this relationship was 0.99. The detection limit of this technology was 0.12 µg/mL CRP. The system used for CRP detection can be extended to also monitor other clinically relevant molecules.

Keywords