Theoretical and Applied Mechanics Letters (Jul 2020)

Ultrasound calibration with ladder phantom at multiple depths for breast biopsy navigation system

  • Jackrit Suthakorn,
  • Narucha Tanaiutchawoot,
  • Cholatip Wiratkapan

Journal volume & issue
Vol. 10, no. 5
pp. 343 – 353

Abstract

Read online

Ultrasound guided breast biopsy navigation system with a graphical user interface and a passive robotic needle holder is developed to increase the performance and reliability of the radiologist. Ultrasound calibration and tool tip calibration are required before using the system. A ladder phantom is developed to be used for ultrasound calibration in real time system with only one ultrasound image required. The passive robotic needle holder structure results in an identity matrix for the makes the rotation matrix; therefore, only translation and scaling are required in the system. This method can be applied to multiple ultrasound depths, which has a relationship at each depth and a relationship to the ultrasound image on the display. The results show high accuracy (<1 mm.) and rapid calibration (5–10 minutes) which is suitable for a real time system like a breast biopsy navigation system based on tests with a breast phantom.