Medicina (Aug 2022)
Comparison of Anatomical Conformity between TomoFix Anatomical Plate and TomoFix Conventional Plate in Open-Wedge High Tibial Osteotomy
Abstract
Background and Objectives: The TomoFix anatomical plate was developed to improve plate position, proximal screw direction, and post-correction tibial contouring. The purpose of this study was to compare postoperative configurations between the TomoFix anatomical plate and the TomoFix conventional plate. It was hypothesized that the new modified plate provides a better fixative coaptation than the conventional plate. Materials and Methods: A total of 116 cases (112 patients) were enrolled in this study from March 2015 to February 2021. Among them, 63 patients underwent surgery using the TomoFix conventional plate, and 53 underwent surgery using the TomoFix anatomical plate. The radiographic outcomes, including the hip–knee–ankle (HKA) angle, medial proximal tibial angle (MPTA), tibial slope, plate angle, proximal screw angles, and plate-to-cortex distance at #1 hole (just below the osteotomy site) were compared between the two groups. Results: Patients with the TomoFix anatomical plate showed similar results in terms of the pre- and postoperative HKA angle, MPTA, and tibial slope. The TomoFix anatomical group showed a significantly greater plate angle (39.2° ± 8.1° vs. 31.7° ± 7.0°, p p Conclusion: The TomoFix anatomical plate showed a more posteromedial plating position, better proximal screw direction to the lateral hinge, and improved post-correction tibial contour compared to the TomoFix conventional plate.
Keywords