Measurement Science Review (Oct 2021)

First Evaluation of the PTN-104 Plethysmographic Sensor for Heart Rate Measurement

  • Szaj Wiesław,
  • Wojnarowska Wiktoria,
  • Pajdo Bogdan

DOI
https://doi.org/10.2478/msr-2021-0017
Journal volume & issue
Vol. 21, no. 5
pp. 117 – 122

Abstract

Read online

The purpose of this study was to examine the accuracy of HR measurements by the PTN-104 sensor in comparison to the fingertip pulse oximeter, which is a photoplethysmographic sensor (PPG). Twelve healthy participants underwent the same protocol during a single visit. Measurements were taken after each participant completed an initial rest period of 5 minutes and after 1-minute of exercising comprising of 30 squats. Each subject had the PTN-104 sensor attached to the index finger and a fingertip pulse oximeter to the opposite one. When examining the data in aggregate, there was a strong correlation between the PTN-104 sensor and PPG for HR (r = 0.988) with a mean bias of −2.55 bpm (95 % LoA +5.0, −10.1). The PTN-104 sensor satisfied validity criteria for HR monitors, however, showed a lower accuracy for measurements at rest, which is surprising. Due to the noticed limitations, this study should be repeated with a larger group of subjects and the PTN-104 sensor should be compared to the gold standard method for measuring HR, which is ECG.