Sensors (Jul 2022)

Improving PMUT Receive Sensitivity via DC Bias and Piezoelectric Composition

  • Christopher Cheng,
  • Travis Peters,
  • Ajay Dangi,
  • Sumit Agrawal,
  • Haoyang Chen,
  • Sri-Rajasekhar Kothapalli,
  • Susan Trolier-McKinstry

DOI
https://doi.org/10.3390/s22155614
Journal volume & issue
Vol. 22, no. 15
p. 5614

Abstract

Read online

The receive sensitivity of lead zirconate titanate (PZT) piezoelectric micromachined ultrasound transducers (PMUTs) was improved by applying a DC bias during operation. The PMUT receive sensitivity is governed by the voltage piezoelectric coefficient, h31,f. With applied DC biases (up to 15 V) on a 2 μm PbZr0.52Ti0.48O3 film, e31,f increased 1.6 times, permittivity decreased by a factor of 0.6, and the voltage coefficient increased by ~2.5 times. For released PMUT devices, the ultrasound receive sensitivity improved by 2.5 times and the photoacoustic signal improved 1.9 times with 15 V applied DC bias. B-mode photoacoustic imaging experiments showed that with DC bias, the PMUT received clearer photoacoustic signals from pencil leads at 4.3 cm, compared to 3.7 cm without DC bias.