Journal of Arrhythmia (Apr 2020)
Use of the new rotating dilator sheath TightRail™ for lead extraction: A bicentric experience
Abstract
Abstract Aim The aim of this study is to assess the safety and efficacy of the new bidirectional rotational mechanical sheath TightRail™ (Spectranetics) for lead extraction. Methods and results This is a bicentric prospective study that included patients who underwent a transvenous lead extraction (TLE) in two Italian centers (San Raffaele Hospital and Humanitas Gavazzeni Hospital). From November 2016 to December 2018, 26 patients underwent a TLE procedure in which the TightRail™ was used. The new TightRail Sub‐C was used in 20 (76%) patients to overcome the fibrosis between the vessel and the first rib. Median age was 69 (IQR 60.7‐79.5) years. The indication for TLE were infection (57.7%) or lead dysfunction (42.3%). A total of 57 leads (range 1‐4), 40 of which using the TightRail (range 1‐4), were extracted. Overall mean implant duration was 98.2.0 ± 66.5 months. Mean age of the lead extracted with the TightRail sheath was 99.1 ± 70.2 months and was higher compared to that of the leads extracted manually (84.4 ± 60.3 months, P = .001). The overall clinical success was 100% and complete procedural success without the use of a snare was achieved in 98.3%. There were no cases of death or major complications and only two minor complications occurred. All patients were event‐free at 6‐month follow‐up. Conclusion This initial experience using the TightRail™ suggests a high safety and efficacy profile for extractions in a wide range of lead age.
Keywords