Scientific Reports (Nov 2021)

The effect of pupil transmittance on axial resolution of reflection phase microscopy

  • Min Gyu Hyeon,
  • Kwanjun Park,
  • Taeseok Daniel Yang,
  • Taedong Kong,
  • Beop-Min Kim,
  • Youngwoon Choi

DOI
https://doi.org/10.1038/s41598-021-02188-0
Journal volume & issue
Vol. 11, no. 1
pp. 1 – 9

Abstract

Read online

Abstract A reflection phase microscope (RPM) can be equipped with the capability of depth selection by employing a gating mechanism. However, it is difficult to achieve an axial resolution close to the diffraction limit in real implementation. Here, we systematically investigated the uneven interference contrast produced by pupil transmittance of the objective lens and found that it was the main cause of the practical limit that prevents the axial resolution from reaching its diffraction limit. Then we modulated the power of illumination light to obtain a uniform interference contrast over the entire pupil. Consequently, we could achieve an axial resolution fairly close to the diffraction limit set by the experimental conditions.