Applied Sciences (Jan 2021)
Method for Horizontal Calibration of Laser-Projection Transnasal Fiberoptic High-Speed Videoendoscopy
Abstract
Objective: Calibrated horizontal measurements (e.g., mm) from endoscopic procedures could be utilized for advancement of evidence-based practice and personalized medicine. However, the size of an object in endoscopic images is not readily calibrated and depends on multiple factors, including the distance between the endoscope and the target surface. Additionally, acquired images may have significant non-linear distortion that would further complicate calibrated measurements. This study used a recently developed in vivo laser-projection fiberoptic laryngoscope and proposes a method for calibrated spatial measurements. Method: A set of circular grids was recorded at multiple working distances. A statistical model was trained that would map from pixel length of the object, the working distance, and the spatial location of the target object into its mm length. Result: A detailed analysis of the performance of the proposed method is presented. The analyses have shown that the accuracy of the proposed method does not depend on the working distance and length of the target object. The estimated average magnitude of error was 0.27 mm, which is three times lower than the existing alternative. Conclusion: The presented method can achieve sub-millimeter accuracy in horizontal measurement. Significance: Evidence-based practice and personalized medicine could significantly benefit from the proposed method. Implications of the findings for other endoscopic procedures are also discussed.
Keywords