PLoS ONE (Jan 2012)

High risk HPV contamination of endocavity vaginal ultrasound probes: an underestimated route of nosocomial infection?

  • Jean-sebastien Casalegno,
  • Karine Le Bail Carval,
  • Daniel Eibach,
  • Marie-Laure Valdeyron,
  • Gery Lamblin,
  • Hervé Jacquemoud,
  • Georges Mellier,
  • Bruno Lina,
  • Pascal Gaucherand,
  • Patrice Mathevet,
  • Yahia Mekki

DOI
https://doi.org/10.1371/journal.pone.0048137
Journal volume & issue
Vol. 7, no. 10
p. e48137

Abstract

Read online

BACKGROUND: Endocavity ultrasound is seen as a harmless procedure and has become a common gynaecological procedure. However without correct disinfection, it may result in nosocomial transmission of genito-urinary pathogens, such as high-risk Human Papillomavirus (HR-HPV). We aimed to evaluate the currently recommended disinfection procedure for covered endocavity ultrasound probes, which consists of "Low Level Disinfection" (LLD) with "quaternary ammonium compounds" containing wipes. METHODS: From May to October 2011 swabs were taken from endovaginal ultrasound probes at the Gynecology Department of the Lyon University Hospital. During the first phase (May-June 2011) samples were taken after the ultrasound examination and after the LLD procedure. In a second phase (July-October 2011) swab samples were collected just before the probe was used. All samples were tested for the presence of human DNA (as a marker for a possible transmission of infectious pathogens from the genital tract) and HPV DNA with the Genomica DNA microarray (35 different HPV genotypes). RESULTS: We collected 217 samples before and 200 samples after the ultrasound examination. The PCR was inhibited in two cases. Human DNA was detected in 36 (18%) post-examination samples and 61 (28%) pre-examination samples. After the ultrasound LLD procedure, 6 (3.0%) samples contained HR-HPV types (16, 31, 2×53 and 58). Similarly, HPV was detected in 6 pre-examination samples (2.7%). Amongst these 4 (1.9%) contained HR-HPV (types 53 and 70). CONCLUSION: Our study reveals that a considerable number of ultrasound probes are contaminated with human and HR-HPV DNA, despite LLD disinfection and probe cover. In all hospitals, where LLD is performed, the endovaginal ultrasound procedure must therefore be considered a source for nosocomial HR-HPV infections. We recommend the stringent use of high-level disinfectants, such as glutaraldehyde or hydrogen peroxide solutions.