GMS Hygiene and Infection Control (Mar 2014)

Effective reprocessing of reusable dispensers for surface disinfection tissues – the devil is in the details

  • Kampf, Günter,
  • Degenhardt, Stina,
  • Lackner, Sibylle,
  • Ostermeyer, Christiane

DOI
https://doi.org/10.3205/dgkh000229
Journal volume & issue
Vol. 9, no. 1
p. Doc09

Abstract

Read online

[english] Background: It has recently been reported that reusable dispensers for surface disinfection tissues may be contaminated, especially with adapted , when products based on surface-active ingredients are used. Fresh solution may quickly become recontaminated if dispensers are not processed adequately. Methods: We evaluated the abilities of six manual and three automatic processes for processing contaminated dispensers to prevent recolonisation of a freshly-prepared disinfectant solution (Mikrobac forte 0.5%). Dispensers were left at room temperature for 28 days. Samples of the disinfectant solution were taken every 7 days and assessed quantitatively for bacterial contamination. Results: All automatic procedures prevented recolonisation of the disinfectant solution when a temperature of 60–70°C was ensured for at least 5 min, with or without the addition of chemical cleaning agents. Manual procedures prevented recontamination of the disinfectant solution when rinsing with hot water or a thorough cleaning step was performed before treating all surfaces with an alcohol-based disinfectant or an oxygen-releaser. Other cleaning and disinfection procedures, including the use of an alcohol-based disinfectant, did not prevent recolonisation.Conclusions: These results indicate that not all processes are effective for processing reusable dispensers for surface-disinfectant tissues, and that a high temperature during the cleaning step or use of a biofilm-active cleaning agent are essential.

Keywords