Journal of Oral Research (Apr 2022)

Comparative evaluation of dental pulp tests in pulpal status diagnostics

  • Roberto E. Campos,
  • Paulo César Freitas Santos Filho,
  • Érice França Resende,
  • Lílian Vieira Oliveira,
  • Gláucia Maria Bovi Ambrosano

DOI
https://doi.org/10.17126/joralres.2022.014
Journal volume & issue
Vol. 11, no. 2
pp. 1 – 11

Abstract

Read online

Introduction: The purpose of this study was to evaluate the effectiveness and compare the accuracy of pulp tests in the diagnosis of teeth pulpal health. Material and Methods: Traumatized (n=71) and non-traumatized (n=71) teeth from 42 patients were evaluated. Each tooth underwent cold, heat, electric and oximetry tests, followed by radiographic examination and calculation of the sensitivity, specificity, PPV, NPV and accuracy. Results: Clinical and radiographic examination showed no alteration for the 71 teeth from the intact contralateral group. From the traumatized group, 29 teeth presented complete endodontic treatment, 17 presented periapical alterations that required endodontic treatment and 25 teeth did not present conclusive radiographic alteration. The cold test showed a significantly higher proportion of correct results, while the electric test showed a significantly lower proportion. The data showed higher accuracy for the cold, followed by oximeter and heat tests, while the electric test presented the lowest accuracy. Cold and oximeter tests proved superior over the electric and heat tests, while the electric test showed better parameters when diagnosing diseased pulp. Conclusion: Combining two pulp tests seems reasonable for improving the pulp diagnoses using both oximeter and cold or oximeter and heat tests to detect healthy pulp; or cold and electric tests to define diseased pulp.

Keywords