Scientific Reports (Feb 2022)

Ultrasound does not activate but can inhibit in vivo mammalian nerves across a wide range of parameters

  • Hongsun Guo,
  • Sarah J. Offutt,
  • Mark Hamilton II,
  • Yohan Kim,
  • Cory D. Gloeckner,
  • Daniel P. Zachs,
  • Jamu K. Alford,
  • Hubert H. Lim

DOI
https://doi.org/10.1038/s41598-022-05226-7
Journal volume & issue
Vol. 12, no. 1
pp. 1 – 14

Abstract

Read online

Abstract Ultrasound (US) has been shown to stimulate brain circuits, however, the ability to excite peripheral nerves with US remains controversial. To the best of our knowledge, there is still no in vivo neural recording study that has applied US stimulation to a nerve isolated from surrounding tissue to confirm direct activation effects. Here, we show that US cannot excite an isolated mammalian sciatic nerve in an in vivo preparation, even at high pressures (relative to levels recommended in the FDA guidance for diagnostic ultrasound) and for a wide range of parameters, including different pulse patterns and center frequencies. US can, however, reliably inhibit nerve activity whereby greater suppression is correlated with increases in nerve temperature. By prohibiting the nerve temperature from increasing during US application, we did not observe suppressive effects. Overall, these findings demonstrate that US can reliably inhibit nerve activity through a thermal mechanism that has potential for various health disorders, though future studies are needed to evaluate the long-term safety of therapeutic ultrasound applications.