Sensors (Jan 2021)

Oxygen Saturation Imaging Using LED-Based Photoacoustic System

  • Rianne Bulsink,
  • Mithun Kuniyil Ajith Singh,
  • Marvin Xavierselvan,
  • Srivalleesha Mallidi,
  • Wiendelt Steenbergen,
  • Kalloor Joseph Francis

DOI
https://doi.org/10.3390/s21010283
Journal volume & issue
Vol. 21, no. 1
p. 283

Abstract

Read online

Oxygen saturation imaging has potential in several preclinical and clinical applications. Dual-wavelength LED array-based photoacoustic oxygen saturation imaging can be an affordable solution in this case. For the translation of this technology, there is a need to improve its accuracy and validate it against ground truth methods. We propose a fluence compensated oxygen saturation imaging method, utilizing structural information from the ultrasound image, and prior knowledge of the optical properties of the tissue with a Monte-Carlo based light propagation model for the dual-wavelength LED array configuration. We then validate the proposed method with oximeter measurements in tissue-mimicking phantoms. Further, we demonstrate in vivo imaging on small animal and a human subject. We conclude that the proposed oxygen saturation imaging can be used to image tissue at a depth of 6–8 mm in both preclinical and clinical applications.