Scientific Reports (Feb 2021)
Reproducibility of 3D scanning in the periorbital region
Abstract
Abstract The reproducibility of scanning in the periorbital region with 3D technology to enable objective evaluations of surgical treatment in the periorbital region was assessed. Facial 3D-scans of 15 volunteers were captured at different time points with a handheld Artec Space Spider structured light scanner. Two scans were made with a one minute interval and repeated after 1 year; for both a natural head position and with the head in a fixation-device. On assessing the area between the eyelashes and eyebrows, the medians of the average deviations between the various cross-sections of the one minute interval 3D-scans ranged from 0.17 to 0.21 mm at baseline, and from 0.10 to 0.11 mm when the minute-interval scanning was repeated one year later. The systematic differences when scanning in a natural head position and fixated position were comparable. The reproducibility of the 3D processing was excellent (intraclass correlation coefficient > 0.9). The repeated scanning deviations (baseline versus one year data) were well within the accepted clinical threshold of 1 mm. Scanning with a hand-held 3D-scanning device (Artec Space Spider) is a promising tool to assess changes in the periorbital region following surgical treatment since the median deviations are well below the clinically accepted 1 mm measuring error, for both the natural head and fixated positions.