IEEE Journal of Translational Engineering in Health and Medicine (Jan 2018)

A Handheld, Colorimetric Optoelectronic Dynamics Analyzer for Measuring Total Ammonia of Biological Samples

  • Nai-Yuan Liu,
  • Pinar Cay-Durgun,
  • Tianmiao Lai,
  • Mark Sprowls,
  • Leslie Thomas,
  • Mary Laura Lind,
  • Erica Forzani

DOI
https://doi.org/10.1109/JTEHM.2018.2840678
Journal volume & issue
Vol. 6
pp. 1 – 10

Abstract

Read online

This paper introduces a wireless, solid-state, portable, and automated device capable of measuring the total ammonia [ammonia (NH3) and ammonium (NH4+)] levels of fluids, including biological samples. This device reliably measures the total ammonia of biological samples (e.g., urine) faster than the current ammonia quantification techniques. Medical professionals typically estimate NH4+ levels using error-prone indirect measurement techniques (i.e., urine anion gap), which are time-consuming and are seldom suitable for periodic measurements. Several instantaneous measurements of total ammonia levels in a patient urine could be utilized as an early warning for both acid-base and/or potassium disturbances. Given the device's operation mechanism, it is able to quantify the total ammonia concentration within a biological sample in only 5 s and can simultaneously transmit data to other devices via Bluetooth. The analytical operation demonstrated high sensitivity, high specificity, fast reversibility, rapid response time, and has enabled the accurate determination of total ammonia concentration in urine samples produced by subjects who had consumed diets of variable protein compositions.

Keywords