Metrology (Jan 2024)

Digital Impedance Bridge for Four-Terminal-Pair AC Resistor Calibration up to 20 kHz

  • Mohamed Ouameur,
  • Renata Vasconcellos,
  • Mohamed Agazar

DOI
https://doi.org/10.3390/metrology4010001
Journal volume & issue
Vol. 4, no. 1
pp. 1 – 14

Abstract

Read online

For this study, a substitution principle-based impedance bridge has been developed to calibrate AC resistors in a four-terminal-pair (4TP) configuration. The calibration is performed in the full complex plane for resistances ranging from 100 mΩ to 400 Ω and frequencies of between 50 Hz and 20 kHz. The automated bridge is based on four resistors associated with two high-impedance stages. The balancing of the bridge is achieved by means of PXI modules. The bridge is automatically balanced via a simplex top-down algorithm. The new bridge is primarily used for the measurement chain of AC standard resistors defined in a 4TP configuration at LNE, which are used for routine customer calibrations. The traceability of LNE’s standard resistors when defined in a 4TP configuration is ensured by a measurement chain from a 1 kΩ reference resistor using the new bridge. The reference resistor was calibrated previously via comparison with a calculable resistor up to 20 kHz. The bridge was validated via comparison with calibration results obtained in 1983 and 2009. For a resistor of 1 Ω at 1 kHz, the uncertainty of the series resistance variation and the phase shift are less than 6 µΩ/Ω (k = 1) and 6 µrad (k = 1), respectively.