Metrology (Jun 2024)

Chromogenic Approach for Oxygen Sensing Using Tapered Coreless Optical Fibre Coated with Methylene Blue

  • Rahul Kumar,
  • Neil Wight

DOI
https://doi.org/10.3390/metrology4020018
Journal volume & issue
Vol. 4, no. 2
pp. 295 – 303

Abstract

Read online

In this paper, a Methylene Blue (MB)-coated tapered coreless (TCL) optical fibre sensor is proposed and experimentally investigated for oxygen sensing in the near-infrared (NIR) wavelength range of 993.5 nm. The effect of TCL diameter and MB sol–gel coating thickness on the sensitivity of the sensor was also investigated. A maximum sensitivity of 0.19 dB/O2% in the oxygen concentration range of 0–37.5% was achieved for a TCL fibre sensor with a 2 µm taper waist diameter and a 0.86 µm MB sol–gel coating thickness, with a response time of 4 min. The sensor provides reproducible results even after 7 days and is shown to be highly selective to oxygen compared to argon and ethanol at the same concentration.

Keywords